Product Reference Guide 6-15
Shift Test (Metric — 2 Kilogram)
The Shift Test checks to ensure that items placed anywhere on the weigh-
ing surface of the scanner/scale ar
e weighed properly. Refer to Figure 6-3
when performing this test.
1. Place and remove, in succession, a 2 kilogram load on the center
o
f each of the four quadrants: 2, 3, 4, and 5, and in the center (1)
of the scanner/scales weighing platform (see
Figure 6-3). Verify
that the Remote Display shows a reading of between 1.998 and
2.002 kg
for each quadrant/center test. Check that the display
returns to 0.000 between each load.
2. After verifying the accuracy of each quadrant and the center of
the
weighing surface, remove all weight from the scale. This com-
pletes the Shift Test (Metric — 2 Kilogram).